Error in measuring the permittivity of conducting media by the broadband method | Izvestiya vuzov. Fizika. 2022. № 7. DOI: 10.17223/00213411/65/7/137

Error in measuring the permittivity of conducting media by the broadband method

The measurement of the complex dielectric permittivity (CDP) of nine calibration liquids with static permittivity from 2.38 (toluene) to 78.5 (water) was carried out. The decision to select the dimensions of the coaxial cell used in the measurement of the CDP of conducting media by the broadband method in the frequency range from hundreds of hertz to 8 GHz is justified. It is shown that the use of a cell with the dimensions of the external and internal conductors of 16 mm and 6.97 mm, respectively, in comparison with a cell with the dimensions of 7 mm and 3.02 mm reduces the frequency of influence of electrode polarization. The relations of the measurement error of the real part of the CDP in the low-frequency range with the loss tangent for liquids with different specific electrical conductivity are found.

Download file
Counter downloads: 29

Keywords

complex dielectric permittivity, specific electrical conductivity, calibration fluids, conducting media, error of dielectric measurements

Authors

NameOrganizationE-mail
Bobrov P.P.Omsk Scientific Center SB RAS (Institute of Radiophysics and Physical Electronics); Omsk State Pedagogical Universitybobrov@omgpu.ru
Rodionova O.V.Omsk State Pedagogical Universityolga_vk07@list.ru
Всего: 2

References

Крошка Е.С. Широкополосная диэлектрическая спектроскопия почв и пористых горных пород: автореф. дис. … канд. физ.-мат. наук: 1.4.3. - Омск, 2022. - 23 с.
Kaatze U., Feldman Y. // Meas. Sci. Technol. - 2006. - V. 17. - P. R17-R35.
Schwing M., Wagner N., Karlovsek J., et al. //j. Geophys. Eng. - 2016. - V. 13. - P. S28-S38.
Bobrov P.P., Repin A.V., Rodionova O.V. // IEEE Trans. Geosci. Remote Sens. - 2015. - V. 53. - No. 5. - P. 2366-2372.
Gorriti A., Slob E. // IEEE Trans. Geosci. Remote Sens. - 2005. - V. 43. - No. 9. - P. 2051-2057.
Keysight E4990A Impedance Analyzer, 20 Hz to 10/20/30/50/120 MHz. - URL: https://www.etesttool.com/keysight-e4990a.html.
Grosse C., Tirado M. //j. Non-Cryst. Solids. - 2002. - V. 305. - P. 386-392.
Бобров П.П., Кондратьева О.В., Репин А.В. // Изв. вузов. Физика. - 2012. - T. 55. - № 8/3. - С. 23-26.
Репин А.В., Беляева Т.А., Родионова О.В., Крошка Е.С. // Изв. вузов. Физика. - 2015. - Т. 58. - № 8/2. - С. 32-35.
Kirichek A., Chassagne C., Ghose R. // Frontiers in Physics. - 2019. - V. 6. - Art. 148.
Xiaobo D., Wang Yu.-H. // Soil Sci. Soc. Am. J. - 2008. - V. 72. - No. 6. - P. 1532-1541.
Peyman A., Gabriel C., Grant E.H. // Bioelectromagnetics. - 2007. - V. 28. - P. 264-274.
Gregory A.P. // Technical report number: MAT 23. - 2012. - URL: https://www.researchgate.net/publication/235800733.
Репин А.В., Родионова О.В., Крошка Е.С. // Изв. вузов. Физика. - 2021. - Т. 64. - № 1. - С. 58-63.
 Error in measuring the permittivity of conducting media by the broadband method | Izvestiya vuzov. Fizika. 2022. № 7. DOI: 10.17223/00213411/65/7/137